  cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput p ower s upply ? 2mopp safety approved fo r medical applications according to iec 60601-1, 3 rd edition ? emc tested for medical use according to iec 60601-1-2, 4 th edition ? quick-connect spring-clamp terminals ? ac 100-240v wide-range input ? width only 39mm ? efficiency up to 95.2%, excelle nt partial load efficiency ? 20% output power reserves ? safe hiccup plus overload mode ? easy fuse breaking due to high overload peak current ? active power factor correction (pfc) ? minimal inrush current surge ? full power between -25c and +60c ? dc-ok relay contact ? 3 years limited warranty 1/27 g eneral d escription the cp10.241-m1 is a din-rail mountable single-phase- input power supply for indu strial use and for use in medical equipment. it provides a stabilized and galvanically separated se lv/pelv output voltage. the specialty of this power supply is the 2mopp (two means of patient protection) safety approval for medical use and the fulfillm ent of the required emc tests for professional heal thcare facility and home healthcare environments. the cp-series is part of the dimension power supply family. the most outstanding features of cp10.241-m1 are the high efficiency, ad vanced inrush current limitation, active pfc and the wide operational temperature range. high immunity to transients a nd power surges as well as low electromagnetic emission, a dc-ok relay contact and a large international appr oval package for a variety of applications makes this un it suitable for nearly every situation. s hort - form d ata output voltage dc 24v adjustment range 24 C 28v factory setting 24.1v output current 12 C 10.3a below +45c ambient 10 C 8.6a at +60c ambient 7.5 C 6.5a at +70c ambient derate linearely between +45c and +70c input voltage ac ac 100 - 240v -15%/ +10% mains frequency 50 - 60hz 6% ac input current 2.15 / 1.13a at 120 / 230vac power factor 0.99 / 0.97 at 120 / 230vac input inrush current 6 / 9a pk at 40c 120 / 230vac input voltage dc dc 110-150v 20% input current dc 2.35a at 110vdc efficiency 93.6 / 95.2% at 120 / 230vac power losses 16.4 / 12.1w at 120 / 230vac hold-up time 37ms temperature range -25c to +70c continuous size (wxhxd) 39x124x117mm without din-rail weight 620g / 1.37lb o rder n umbers power supply cp10.241-m1 medical approved device accessory zm12.side side mount bracket m arkings ind. cont. e q. ul 508, planned ul 60950-1, planned emc, lvd medical 2mopp - iec 60601-1 may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput i ndex page page 1. intended use .......................................................3 2. installation requirements ...................................3 3. ac-input ...............................................................4 4. dc-input ...............................................................5 5. input inrush current ...........................................6 6. output .................................................................7 7. hold-up time .......................................................8 8. dc-ok relay contact ..........................................9 9. efficiency and power losses ..............................10 10. lifetime expectancy and mtbf .........................11 11. functional diagram ...........................................11 12. terminals and wiring ........................................12 13. front side and user elements ...........................13 14. emc C according to generic standards ...........14 15. emc C according to medical standards ...........15 16. environment ......................................................16 17. safety features ..................................................17 18. protection features ...........................................18 19. dielectric strength ............................................18 20. approvals ...........................................................19 21. rohs, reach and othe r fulfilled standards ...19 22. physical dimensions and weight ..................... 20 23. accessories ........................................................ 21 23.1. zm12.side - side mounting bracket ........21 23.2. yr20.242 - redundancy module ..............22 23.3. yr20.246 - redundancy module with automated load sharing ....................................22 24. application notes ............................................. 23 24.1. peak current capability ...........................23 24.2. back-feeding loads ..................................24 24.3. external input protection .........................24 24.4. output circuit breakers ............................24 24.5. series operation .......................................25 24.6. parallel use to increase output power ....25 24.7. parallel use for redundancy ....................25 24.8. inductive and capacitive loads ................25 24.9. charging of batteries ...............................25 24.10. operation on two phases ........................26 24.11. use in a tightly sealed enclosure ............26 24.12. mounting orientations ............................27 the information given in this document is correct to the best of our kn owledge and experience at the time of publication. if not expressly agreed othe rwise, this information does not represent a warranty in the legal sense of the word. as the state of our knowledge an d experience is constantly changing, the information in this data sheet is subject to revision. we therefore ki ndly ask you to always use the latest issue of this document (available under www.pulspower.com). no part of this document may be reproduced or utilized in any form without our prior permission in writing. t erminology and a breviations pe and symbol pe is the abbreviation for p rotective e arth and has the same meaning as the symbol . earth, ground this document uses the term earth which is the same as the u.s. term ground. t.b.d. to be defined, value or de scription will follow later. ac a lternating current dc d irect c urrent ac 230v a figure displayed with the ac or dc before the value represents a nominal voltage with standard tolerances (usually 15%) included. e.g.: dc 12v describes a 12v battery disregardi ng whether it is full (13.7v) or flat (10v) 230vac a figure with the unit (vac) at the end is a momentary figure without any additional tolerances included. 50hz vs. 60hz as long as not otherwise stated, ac 100v a nd ac 230v parameters ar e valid at 50hz mains frequency. ac 120v parameters are valid for 60hz mains frequency. may a key word indicating flexibility of choice with no im plied preference. shall a key word indicating a mandatory requirement. should a key word indicating flexibility of choi ce with a strongly pref erred implementation. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 2/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 1. i ntended u se this device is intended for use in medi cal equipment as well as fo r general use such as in industrial control, office, communication, and instrumentation equipment. do not use this device in equipment, where malfunction ma y cause severe personal injury or threaten human life. 2. i nstallation r equirements warning risk of electrical shock, fi re, personal injury or death. - turn power off before working on the devi ce. protect against inad vertent re-powering. - do not open, modify or repair the unit. - use caution to prevent any foreign objects from enteri ng the housing. - do not use in wet locations or in areas wh ere moisture or condens ation can be expected. - do not touch during power-on, and immediatel y after power-off. hot surfaces may cause burns. this device may only be installed and put into opera tion by qualified personnel. this device does not contain serviceable parts. the tripping of an internal fuse is caused by an internal defect. if damage or malfunction should occur dur ing installation or operation, immediatel y turn power off and send the device to the factory for inspection. install the device in an enclosure providing protecti on against electrical, mechan ical and fire hazards. install the device onto a din-rail accord ing to en 60715 with the input termina ls on the bottom of the device. for other orientations see de-rating requirements in chapter 24.12. make sure that the wiring is correct by following all local and national codes. use appropriate copper cables that are designed for a minimum operating te mperature specified in chapter 12 . do not use the device in pollution degree 3 environments without additional protection or in applications where a degree of protection better than ip20 is required. a disconnecting means shall be provided for the output of the device when us ed in applications according to csa c22.2 no 107.1-01. the device is designed as class of pr otection i equipment according to ie c 61140. a pe (protective earth) connection is required. the device is designed for convection co oling and does not require an external fan. do not obstruct airflow and do not cover ventilation grid (e.g. cabl e conduits) by more than 15%! keep the following minimum installation cl earances when the device is permanent ly loaded with more than 50% of the nominal current: 40mm on top, 20mm on the bottom, 5mm left and right side . increase the 5mm to 15mm in case the adjacent device is a heat source. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 3/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 3. ac-i nput ac input nom. ac 100-240v suitable for tn-, tt- and it mains networks ac input range min. 85-264vac continuous operation min. 264-300vac for max. 500ms allowed voltage l or n to earth max. 300vac continuous, iec 62477-1 input frequency nom. 50C60hz 6% turn-on voltage typ. 80vac steady-state value, see fig. 3-1 shut-down voltage typ. 70vac steady-state value, see fig. 3-1 typ. 55vac dynamic value (max. 250ms) external input protection see recommendations in chapter 24.3 . ac 100v ac 120v ac 230v input current typ. 2.60a 2. 15a 1.13a at 24v, 10a, see fig. 3-3 power factor *) typ. 0.99 0.99 0.97 at 24v, 10a, see fig. 3-4 crest factor **) typ. 1.5 1.5 1.65 at 24v, 10a start-up delay typ. 300ms 290ms 240ms see fig. 3-2 rise time typ. 30ms 30ms 30ms at 24v, 10a const. current load, 0mf load capacitance, see fig. 3-2 typ. 75ms 75ms 75ms at 24v, 10a const. current load, 10mf load capacitance,, see fig. 3-2 turn-on overshoot max. 200mv 200mv 200mv see fig. 3-2 external input protection see recommendations in chapter 24.3 . *) the power factor is the ratio of the true (or real) power to the apparent power in an ac circuit. **) the crest factor is the mathematical ratio of the peak value to rms value of the input current waveform. fig. 3-1 input voltage range fig. 3-2 turn-on behavior, definitions turn-on 85v rated input range max. 500ms v in p out 300vac 264v shut-down start-up delay rise time overshoot - 5% output voltage input voltage may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 4/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput fig. 3-3 input current vs. output current at 24v output voltage fig. 3-4 power factor vs. output current at 24v output voltage 12a 123456789 0 0.5 1.0 1.5 2.0 2.5 3a input current, typ. 10 11 a) 100vac b) 120vac c) 230vac b a c output current power factor, typ. 123456789 12a 0.75 0.8 0.85 0.9 0.95 1.0 10 11 output current (a) 100vac, (b) 120vac, (c) 230vac (a) (b) (c) 4. dc-i nput dc input nom. dc 110-150v 20% dc input range min. 88-180vdc continuous operation dc input current typ. 2.35a at 110vdc, at 24v, 10a allowed voltage l/n to earth max. 375vdc continuous, iec 62477-1 turn-on voltage typ. 80vdc steady state value shut-down voltage typ. 70vdc steady state value typ. 55vdc dynamic value (max. 250ms) fig. 4-1 wiring for dc input instructions for dc use: + - load l pe + - power supply ac dc battery n a) use a battery or a similar dc source. a supply from the intermediate dc-bus of a frequency converter is not recommended and can cause a malfunction or damage the unit. b) connect +pole to l and Cpole to n. c) connect the pe terminal to an earth wire or to the machine ground. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 5/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 5. i nput i nrush c urrent an active inrush limitation circuit (ntc s, which are bypassed by a relay contact) limits the input inrush current after turn-on of the input voltage. the charging current into emi suppression capacitors is disregarded in th e first microseconds after switch-on. ac 100v ac 120v ac 230v inrush current max. 11a peak 7a peak 11a peak at 40c, cold start typ. 9a peak 6a peak 6a peak at 25c, cold start typ. 9a peak 6a peak 9a peak at 40c, cold start inrush energy max. 0.1a2s 0.1a2s 0.4a2s at 40c, cold start fig. 5-1 typical turn-on behaviour at nominal load, 120vac input and 25c ambient fig. 5-2 typical turn-on behaviour at nominal load, 230vac input and 25c ambient 50ms/div input current 2a/div input voltage 250v/div output voltage 20v/div 6a 50ms/div input current 2a/div input voltage 500v/div output voltage 20v/div 6a may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 6/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 6. o utput output voltage nom. 24v adjustment range min. 24-28v guaranteed max. 30.0v ****) at clockwise end position of potentiometer factory settings typ. 24.1v 0.2 %, at full load and cold unit line regulation max. 10mv 85-300vac load regulation max. 50mv static value, 0a ? 10a; see fig. 6-1 ripple and noise voltage max. 50mvpp 20hz to 20mhz, 50ohm output current nom. 10a at 24v , ambient temperature <60c, see fig. 6-1 nom. 12a *) at 24v, ambient temperature <45c, see fig. 16-1 nom. 7.5a at 24v and 70c ambient temperature, see fig. 16-1 nom. 8.6a at 28v, ambien t temperature <60c, see fig. 6-1 nom. 10.3a *) at 28v, ambient temperature <45c, see fig. 16-1 nom. 6.45a at 28v and 70c ambient temperature, see fig. 16-1 typ. 30a up to at least 12ms *****) , output voltage stays above 20v, see fig. 6-2 and fig. 24-3 , this peak current is available once every five seconds (hardware controlled). output power nom. 240w continuously available nom. 288w *) power boost *) overload behaviour cont. current output voltage > 13vdc, see fig. 6-1 hiccup plus mode **) output voltage < 13vdc, see fig. 6-1 short-circuit cu rrent min. 12.5a ***) load impedance 45mohm, see fig. 6-3 max. 15.5a ***) load impedance 45mohm, see fig. 6-3 max. 5a average (r.m.s.) current, load impedance 50mohm, see fig. 6-3 min. 28a up to 12ms, load impedance 45mohm, see fig. 6-2 typ. 30.5a up to 12ms, load impedance 45mohm, see fig. 6-2 output capacitance typ. 4 400f included inside the power supply *) power boost this power/ current is continuously allowed up to an ambient temperature of 45c. above 45c, do not use this power/ current longer than a duty cy cle of 10% and/ or not longer than 1 minute every 10 minutes. **) hiccup plus mode at heavy overloads (when output voltage falls below 13v), the power supply delivers continuous output current for 2s. after thi s, the output is switched off for approx. 18s before a new start attempt is automatically performed. this cycle is repeated as long as the overload exists. if the overload has been cleared, the device will operate normally. see fig. 6-3 ***) discharge current of output capacitors is not included. ****) this is the maximum output voltage which can occur at the clockwise end position of the potentiometer due to tolerances. it is not a guaranteed value which can be achieved. the typical value is about 28.5v. *****) reduced pulse length for ac 100v mains. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 7/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput fig. 6-1 output voltage vs. output current, typ. fig. 6-2 dynamic output current capability, typ. output voltage 0 0 12.5 4 8 12 28v 16 20 24 20a 7.5 2.5 5 10 15 17.5 adjustment range output current continuous current factory setting hiccup plus mode output voltage (dynamic behavior, < 12ms) 0 0 4 8 12 28v 16 20 24 50a 20 10 30 40 515253545 adjustment range output current fig. 6-3 short-circuit on output, hiccup plus mode, typ. output current 0 14a 18s 18s 18s 2s 2s 2s t short -circuit normal operation normal operation 7. h old - up t ime ac 100v ac 120v ac 230v hold-up time typ. 73ms 73ms 73ms at 24v, 5a, see fig. 7-1 min. 55ms 55ms 55ms at 24v, 5a, see fig. 7-1 typ. 37ms 37ms 37ms at 24v, 10a, see fig. 7-1 min. 28ms 28ms 28ms at 24v, 10a, see fig. 7-1 fig. 7-1 hold-up time vs. input voltage fig. 7-2 shut-down behavior, definitions 0 10 20 30 40 80ms 90 120 155 190 230vac input voltage 50 60 hold-up time 70 a b c d a) 24v 5a typ. b) 24v 5a min. c) 24v 10a typ. d) 24v 10a min. - 5% hold-up time zero transition output voltage input voltage may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 8/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 8. dc-ok r elay c ontact this feature monitors the output voltage on the output terminals of a running power supply. contact closes as soon as the output voltage reaches typ. 90% of the adjusted output voltage level. contact opens as soon as the output voltage dips more than 10% belo w the adjusted output voltage. short dips will be extended to a signal length of 100ms. dips shorter th an 1ms will be ignored. contact ratings max. 60vdc 0.3a, 30vdc 1a, 30vac 0.5a resistive load min. 1ma at 5vdc min. permissible load isolation voltage see dielec tric strength table in section 18. fig. 8-1 dc-ok relay contact behavior 100ms 0.9* v adj < 1ms 10% open v out = v adj open closed closed > 1ms may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 9/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 9. e fficiency and p ower l osses ac 100v ac 120v ac 230v efficiency typ. 92.9% 93.6% 95.2% at 24v, 10a typ. 92.5% 93.4% 95.1% at 24v, 12a (power boost) average efficiency *) typ. 92.5% 93.0% 94.3% 25% at 2.5a, 25% at 5a, 25% at 7.5a. 25% at 10a power losses typ. 2.5w 2.1w 1.8w at 24v, 0a typ. 9.8w 8.9w 7.1w at 24v, 5a typ. 18.3w 16.4w 12.1w at 24v, 10a typ. 23.4w 21.7w 14.8w at 24v, 12a (power boost) *) the average efficiency is an assumption for a typical application where the power supply is loaded with 25% of the nominal load for 25% of the time, 50% of the nominal load for another 25% of the time, 75% of the nominal load for another 25% of the time and with 100% of the nominal load for the rest of the time. fig. 9-1 efficiency vs. output current at 24v, typ. fig. 9-2 losses vs. output current at 24v, typ. efficiency 234567 910 12a 90 91 92 93 94 95 96% output current 811 ( a ) ( b ) ( c ) (a) 100vac (b) 120vac (c) 230vac power losses 01 34 67 12a 5 0 10 15 20 30w 910 25 2 5 8 output current 11 ( a ) ( b ) ( c ) (a) 100vac (b) 120vac (c) 230vac fig. 9-3 efficiency vs. input voltage at 24v, 10a, typ. fig. 9-4 losses vs. input voltage at 24v, 10a, typ. efficiency 120 180 230 264vac 90 91 92 93 input voltage 94 95 96% 100 power losses 120 180 230 264vac 10 12 14 16 input voltage 18 20 22w 100 may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 10/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 10. l ifetime e xpectancy and mtbf ac 100v ac 120v ac 230v lifetime expectancy *) 128 000h 141 000h *) 176 000h *) at 24v, 5a and 40c 61 000h 75 000h 120 000h at 24v, 10a and 40c 47 000h 59 000h 101 000h at 24v, 12a and 40c 363 000h *) 399 000h *) 499 000h *) at 24v, 5a and 25c 173 000h *) 211 000h *) 338 000h *) at 24v, 10a and 25c 132 000h *) 166 000h *) 286 000h *) at 24v, 12a and 25c mtbf **) sn 29500, iec 61709 550 000h 560 000h 661 000h at 24v, 10a and 40c 1 003 000h 1 017 000h 1 176 000h at 24v, 10a and 25c mtbf **) mil hdbk 217f 188 000h 188 000h 213 000h at 24v, 10a and 40c; ground benign gb40 252 000h 252 000h 290 000h at 24v, 10a and 25c; ground benign gb25 40 000h 40 000h 47 000h at 24v, 10a and 40c; ground fixed gf40 51 000h 51 000h 61 000h at 24v, 10a and 25c; ground fixed gf25 *) the lifetime expectancy shown in the table indicates the minimum operating hours (service life) and is determined by the lifetime expectancy of the built-in electrolytic capacitors. lifetime expectancy is specified in operational hours and is calculated acc ording to the capacitors manufacturer specification. the manufacturer of the electrolytic capacitors only guarantees a maximum life of up to 15 years (131 400h). any number exceeding this value is a calculated theoretical lifetime which can be used to compare devices. **) mtbf stands for m ean t ime b etween f ailure, which is calculated according to statistical device failures, and indicates reliability of a device. it is the statistical representation of the likelihood of a unit to fail and does not necessarily represent the life of a product. the mtbf figure is a statistical representation of the likelihood of a device to fail. a mtbf figure of e.g. 1 000 000h means that statistically one unit will fail ev ery 100 hours if 10 000 units ar e installed in the field. how ever, it cannot be determined i f the failed unit has been running for 50 000h or only for 100h. 11. f unctional d iagram fig. 11-1 functional diagram + + - - output over- voltage protection pfc converter output voltage regulator power converter output filter output voltage monitor output power manager temper- ature shut- down input fuse input filter input rectifier inrush current limiter v out l n dc-ok contact dc-ok led dc-ok relay - may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 11/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 12. t erminals and w iring the terminals are ip20 finger sa fe constructed and suitable fo r field- and factory wiring. input and output dc-ok-signal type quick-connect spri ng-clamp terminals push-in terminals solid wire max. 6mm 2 max. 1.5mm 2 stranded wire max. 4mm 2 max. 1.5mm 2 american wire gauge max. awg 20-10 max. awg 28-16 wire diameter (including ferru les) max. 2.8mm max. 1.6mm wire stripping length 10mm / 0.4inch 7mm / 0.28inch screwdriver not app licable not required recommended tightening torque no t applicable not applicable instructions: a) use appropriate copper cables that are designed for minimum operating temperatures of: 90c for ambient between 45c and 60c minimum 105c for ambient up to 70c minimum. b) follow national installation c odes and installation regulations! c) ensure that all strands of a strande d wire enter the terminal connection! d) do not use the unit without pe connection. e) unused terminal compartments should be securely tightened. f) ferrules are allowed. daisy chaining: daisy chaining (jumping from one power s upply output to the next) is allowed as long as the average output current through one terminal pin does not exceed 25a. if the current is higher, use a separate distribution terminal block as shown in fig. 12-2 . fig. 12-1 daisy chaining of outputs fig. 12-2 using distribution terminals load + - power supply + + - - output power supply + + - - output max 25a! continuous distribution terminals load + - power supply + + - - output power supply + + - - output may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 12/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 13. f ront s ide and u ser e lements fig. 13-1 a input terminals quick-connect spri ng-clamp terminals n, l line input pe (protective earth) input b output terminals (two identical + poles an d three identical - poles) quick-connect spri ng-clamp terminals + positive output C nega tive (return) output c output voltage potentiometer open the flap to adjust the output voltage. factory set: 24.1v d dc-ok led (green) on, when the output voltage is >90% of the adjusted output voltage e dc-ok relay contact (push-in terminals) monitors the output voltage of the running power supply. see chapter 8 for details. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 13/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 14. emc C a ccording to g eneric s tandards in regards to emc, the power supply is designed for applications in medical applic ations, industrial environment as well as in residentia l, commercial and light industry environment. emc immunity according to generic standards: en 61000-6-1 and en 61000-6-2 electrostatic disc harge en 61000-4-2 contact discharge air discharge 8kv 15kv criterion a criterion a electromagnetic rf field en 61000-4-3 80mhz-2.7ghz 20v/m criterion a fast transients (burst) en 61000-4-4 input lines output lines dc-ok signal (coupling clamp) 4kv 2kv 2kv criterion a criterion a criterion a surge voltage on input en 61000-4-5 l ? n l ? pe, n ? pe 2kv 4kv criterion a criterion a surge voltage on output en 61000-4-5 + ? - + / - ? pe 1kv 2kv criterion a criterion a surge voltage on signals en 61000-4-5 dc-ok signal ? pe 1kv criterion a conducted disturbance en 61000-4- 6 0.15-80mhz 20v criterion a mains voltage dips en 61000-4-11 0% of 100vac 40% of 100vac 70% of 100vac 0% of 200vac 40% of 200vac 70% of 200vac 0vac, 20ms 40vac, 200ms 70vac, 500ms 0vac, 20ms 80vac, 200ms 140vac, 500ms criterion a criterion c criterion a criterion a criterion a criterion a voltage interruptions en 61000-4-11 0% of 200vac 0vac, 5000ms criterion c voltage sags semi f47 0706 dips on the in put voltage according to semi f47 standard 80% of 120vac (96vac) 70% of 120vac (84vac) 50% of 120vac (60vac) 1000ms 500ms 200ms criterion a criterion a criterion a powerful transients vde 0160 over entire load range 750v, 0.3ms criterion a criterions: a: power supply shows normal operation behavior within the defined limits. c: temporary loss of function is possible. power supply may shut-down and restarts by itself. no damage or hazards for the powe r supply will occur. emc emission according to generic standards: en 61000-6-3 and en 61000-6-4 conducted emission input lines en 55011, en 55015, en 55022, fcc part 15, cispr 11, cispr 22 class b for ac input voltages and class a for dc input voltages conducted emission output lines **) iec/cispr 16-1-2, iec/cispr 16-2-1 12db higher than aver age limits for dc power port according to en 61000-6-3 **) radiated emission en 55011, en 55022 class b harmonic input current en 61000-3-2 class a fulfilled between 0a and 12a load class c fulfilled between 6a and 12a load voltage fluctuations, flic ker en 61000-3-3 fulfilled *) this device complies with fcc part 15 rules. operation is subjected to fo llowing two conditions: (1) this device may not cause ha rmful interference, and (2) this device must accept any interferenc e received, including interference that may cause undesired operation. *) tested with constant current loads, non pulsing **) for information only, not mandatory for en 61000-6-3 or en 61000-6-4 restrictions apply only for applications in residential, commercial and light-industrial environments, where local dc power ne tworks according to en 61000-6-3 are involved. no restricti ons for all kinds of industrial applications. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 14/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 15. emc C a ccording to m edical s tandards emc immunity according to medical standard: en 60601-1-2 electrostatic disc harge en 61000-4-2 contact discharge air discharge air discharge dc ok signal 8kv 15kv 8kv criterion a criterion a criterion a electromagnetic rf field en 61000-4-3 80mhz-2.7ghz 10v/m criterion a en 61000-4-3, en 60601-1-2 table 9 385mhz-5.78ghz 9-28v/m criterion a fast transients (burst) en 61000-4-4 input lines output lines dc-ok signal (coupling clamp) 2kv 1kv 1kv criterion a criterion a criterion a surge voltage on input en 61000-4-5 l ? n l ? pe, n ? pe 1kv 2kv criterion a criterion a conducted disturbance en 61000-4- 6 0.15-80mhz 6v criterion a rated power frequency magnetic fields en 61000-4-8 frequency 50hz frequency 60hz 30a/m 30a/m criterion a criterion a mains voltage dips en 61000-4-11 0% of 100vac 0% of 100vac 70% of 100vac 0% of 240vac 0% of 240vac 70% of 240vac 0vac, 10ms 0vac, 20ms 70vac, 500ms 0vac, 10ms 0vac, 20ms 168vac, 500ms criterion a criterion a criterion a criterion a criterion a criterion a voltage interruptions en 61000-4-11 0% of 100vac 0% of 240vac 0vac, 5000ms 0vac, 5000ms criterion b criterion b criterions: a: power supply shows normal operation behavior within the defined limits. b: output voltage out of range or switches off. dc-ok signal might trigger. restores automatically after the test. emc emission according to generic standard: en 60601-1-2 conducted emission input lines en 55011, cispr 11 class b for ac input voltages and class a for dc input voltages radiated emission en 55011, cispr 11 class b harmonic input current en 61000-3-2 class a fulfilled between 0a and 12a load class c fulfilled between 6a and 12a load voltage fluctuations, flic ker en 61000-3-3 fulfilled *) switching frequencies the power supply has three converters with tw o different switching frequencies included. switching frequency 1 110khz pfc converter, input voltage and out put power dependent switching frequency 2 110khz to 140khz ma in converter, output power dependent switching frequency 3 60kh z auxiliary converter may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 15/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 16. e nvironment operational temperature *) -25c to +70c (-13f to 158f) reduce output power according fig. 16-1 non-operational temperature -40c to +85c (-40f to 185f) for storage and transportation output de-rating 3.2w/c 6w/c 45c to 60c (113f to 140f) 60c to 70c (140f to 158f) humidity **) 5 to 95% r.h. for operati on, storage and transportation according to iec 60068-2-30 atmospheric pressure 106-70kpa for ope ration, storage and transportation vibration sinusoidal 2-17.8h z: 1.6mm; 17.8-500hz: 2g ***) 2 hours / axis ***) iec 60068-2-6 shock 30g 6ms, 20g 11ms ***) 3 bumps / direction, 18 bumps in total iec 60068-2-27 altitude 0 to 3000m (0 to 9 840ft) for medical applications 0 to 2000m (0 to 6 560ft) for all other applications 2000 to 6000m (6 560 to 20 000ft) reduce output power or ambient temperature, see fig. 16-2 . altitude de-rating 15w/1000m or 5c/1000m above 2000m (6500ft), see fig. 16-2 over-voltage category iii altitudes up to 2000m according to iec/en 62477-1, ii for altitudes from 2000m to 6000m degree of pollution 2 according to iec/en 62477-1, not conductive labs compatibility the unit does not re lease any silicone or other labs-criti cal substances and is suitable for use in paint shops. *) operational temperature is the same as the ambient or surrounding temperature and is defined as the air temperature 2cm bel ow the unit. **) do not energize while condensation is present ***) tested in combination with din-rails according to en 60715 with a height of 15mm and a thickness of 1.3mm and standard orientation. fig. 16-1 output current vs. ambient temp. fig. 16-2 output current vs. altitude allowed output current at 24v 0 -25 0 20 40 70c 2a 4a 6a 8a 10a 12a 60 a ambient temperature a... 85 to 264vac, continuous b... short term b 0 0 2000m 4000m 6000m 2a 4a 6a 8a 10a 12a allowed output current at 24v altitude b c a . . . t a m b < 6 0 c b . . . t a m b < 5 0 c c . . . t a m b < 4 0 c d . . . s h o r t t e r m a d may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 16/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 17. s afety f eatures input / output separation 1) selv iec/en 60950-1 pelv iec/en 60204-1, iec/en 62477-1, iec 60364-4-41 double or reinforced insulation transformers safety isolating transformers acc. iec/en 61558-2-16 safety isolating transformer s corresponding to part 2-6 of the iec/en 61558 class of protection i pe (protect ive earth) connection required isolation resistance min. 500mohm input to output, measured with 500vdc pe resistance max. 0.1ohm pe terminal to enclosure earth leakage current, industrial typ. 0.06ma / 0.17ma 100vac, 50h z, tn-,tt-mains / it-mains typ. 0.09ma / 0.24ma 120vac, 60h z, tn-,tt-mains / it-mains typ. 0.15ma / 0.42ma 230vac, 50h z, tn-,tt-mains / it-mains max. 0.08ma / 0.20ma 110vac, 50h z, tn-,tt-mains / it-mains max. 0.11ma / 0.30ma 132vac, 60h z, tn-,tt-mains / it-mains max. 0.20ma / 0.55ma 264vac, 50h z, tn-,tt-mains / it-mains earth leakage current 2) 3) , medical typ. 0.21ma 264vac, 60hz, normal condition max. 0.24ma 264vac, 60hz, normal condition typ. 0.41ma 264vac, 60hz, single fault condition max. 0.47ma 264vac, 60hz, single fault condition touch current 2) 3) , medical, mains to enclosure typ. 0.001ma 264vac, 60hz, normal condition max. 0.001ma 264vac, 60hz, normal condition typ. 0.21ma 264vac, 60hz, single fault condition max. 0.24ma 264vac, 60hz, single fault condition touch current 2) 3) , medical, mains to output gnd (-) pole typ. 0. 009ma 264vac, 60hz, normal condition max. 0.015ma 264vac, 60hz, normal condition typ. 0.13ma 264vac, 60hz, single fault condition max. 0.15ma 264vac, 60hz, single fault condition touch current 2) 3) , medical, mains to output plus (+) pole typ. 0.008ma 264vac, 60hz, normal condition max. 0.013ma 264vac, 60hz, normal condition typ. 0.14ma 264vac, 60hz, single fault condition max. 0.16ma 264vac, 60hz, single fault condition 1) double or reinforced insulation 2) after humidity preconditioning treatment 3) no classification according class b, bf and cf since the power supply is not suitable for application parts with direct pati ent contact. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 17/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 18. p rotection f eatures output protection electronically protected against overload , no-load and short-circuits *) output over-voltage pr otection typ. 30.5vdc max. 32vdc in case of an internal pow er supply defect, a redundant circuit limits the maximum output voltage. the output shuts down and automatically attempts to restart. degree of protecti on ip 20 en/iec 60529 for use in a controlled environment according to csa 22.2 no 107.1-01. penetration protection > 4mm e.g. screws, small parts over-temperature protection yes output shut-down with automatic restart input transient protection mo v (metal oxide varistor) internal input fuse included (1 fuse in l-line) not user replaceable *) in case of a protection event, audible noise may occur. 19. d ielectric s trength the output voltage is floating and has no ohmic connection to the ground. type and factory tests are conducted by the manufacturer. field tests may be c onducted in the field using the appropr iate test equipment which applies the voltage with a slow ramp (2s up and 2s down). connect all input-terminals togeth er as well as all ou tput poles before conducting the test. when testing, set the cut-off current settings to the value in the table below. fig. 19-1 dielectric strength a b c d type test 60s 3000vac 4500vac 1500vac 500vac factory test 5s 3000vac 4000vac 1000vac 500vac field test 5s 2700vac 3500vac 800vac 500vac cut-off current setting > 10ma > 5ma > 5ma > 1ma a d c b b *) l input dc-ok 13 14 earth, pe output + - n b*) when testing input to dc-ok ensure that the max. voltage between dc-ok and the output is not exceeded (column d). we recomm end connecting dc-ok pins and the output pins together when performing the test. insulation safety ratings input to ground mopp, moop input to output 2mopp, 2moop output to ground mopp, moop may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 18/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 20. a pprovals ec declaration of conformity the ce mark indicates conformance with the - emc directive and the - low-voltage directive (lvd) iec 60601-1 3 nd edition cb scheme, medical electrical equipment C part 1: general requirements for basic safety and essential performance 2x mopp and 2x moop (except risk assessment) iec 60950-1 2 nd edition planned cb scheme, information technology equipment ul 508 planned ind. cont. e q. listed for use as industri al control equipment; u.s.a. (ul 508) and canada (c22.2 no. 107-1-01); e-file: e198865 ul 60950-1 2 nd edition planned recognized for use as information technology equipment, level 5; u.s.a. (ul 60950-1) and canada (c22.2 no. 60950-1); e-file: e137006 applicable for altitudes up to 2000m. 21. r o hs, reach and o ther f ulfilled s tandards rohs directive directive 2011/65/eu of the european parliament and the council of june 8 th , 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment. reach directive directive 1907/2006/eu of the european parliament and the council of june 1 st , 2007 regarding the registration, evaluation, authorisation a nd restriction of chemicals (reach) en 60601-1-2 emc medical medical electrical equipment - part 1-2: general requirements for basic safety and essent ial performance - collateral standard: electromagnetic comp atibility - requirements and tests notes: - clause 4.1 (rma) is exempted - to maintain basic safety in regards to emc check pe connection every year. units wi th mechanical defects or corrosive surfaces should no longer be used. may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 19/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 22. p hysical d imensions and w eight width 39mm 1.54 height 124mm 4.88 depth 117mm 4.61 the din-rail height must be added to the unit depth to calculate the total required installation depth. weight 620g / 1.37lb din-rail use 35mm din-rails according to en 60715 or en 50022 with a height of 7.5 or 15mm. housing material body: aluminium alloy cover: zinc-plated steel installation clearances see chapter 2 fig. 22-1 front view fig. 22-2 side view may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 20/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 23. a ccessories 23.1. zm12.side - s ide m ounting b racket this bracket is used to mount the device si deways with or without utilizing a din-rail. the two aluminum brackets and the black pl astic slider of the unit have to be de tached, so that the steel brackets can be mounted. for sideway din-rail mounting, the remo ved aluminum brackets and the black pl astic slider need to be mounted on the steel bracket. fig. 23-1 side mounting without din-rail brackets fig. 23-2 side mounting with din-rail brackets fig. 23-3 mounting dimensions side mounting bracket may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 21/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 23.2. yr20.242 - r edundancy m odule the yr20.242 is equipped with two input channels, which are indivi dually decoupled by utilizing mosfet technology. using mosfetss instead of diodes reduces the h eat generation and the voltage drop between input and output. the yr20.242 does not requi re an additional auxiliary voltage and is self-powered even in ca se of a short circuit across the output. due to the low power losses, the unit is very sl ender and only requires 32mm width on the din-rail. the yr20.242 can be used for n+1 and 1+1 redundancy systems. further information and wiring configur ations can be found in chapter 24.7 . 23.3. yr20.246 - r edundancy m odule with a utomated l oad s haring the yr20.246 is equipped with two input channels, which are indivi dually decoupled by utilizing mosfet technology. using mosfetss instead of diodes reduces the h eat generation and the voltage drop between input and output. the yr20.246 does not requi re an additional auxiliary voltage and is self-powered even in ca se of a short circuit across the output. 22/27 due to the low power losses, the unit is very sl ender and only requires 32mm width on the din-rail. the yr20.246 is optimized fo r 1+1 redundancy systems. compared to the yr20.242, the yr20. 246 is featured with an automa ted load sharing between the connected power supplies. the yr20.246 monitors the f unction of the redundancy circuitry and provides a signal in case of too high of output current, whic h could prevent redundancy, if one power supply fails. further information and wiring configur ations can be found in chapter 24.7 . may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 24. a pplication n otes 24.1. p eak c urrent c apability the unit can deliver peak currents (up to several milliseconds) which are higher than the sp ecified short term currents. this helps to start current demanding loads. solenoids, contactors and pneumatic modules often have a steady state coil and a pick-up coil. the inrush current demand of the pick-up coil is several times higher than the steady-state current and usually exceeds the nominal output current (including the powerboo st). the same situation applies when starting a capacitive load. the peak current capability also ensures the safe operati on of subsequent circuit breakers of load circ uits. the load branches are often individually protected with circuit breakers or fuses. in ca se of a short or an overload in one branch circuit, the fuse or circuit breaker need a certain amount of over-current to open in a timely manner. this avoids voltage loss in adjacent circuits. the extra current (peak current) is supplie d by the power converter and the built -in large sized output capacitors of the power supply. the capacito rs get discharged during such an event, which causes a vo ltage dip on the output. the following two examples show typical voltage dips: fig. 24-1 peak load with 2x the nominal current for 50ms, typ. fig. 24-2 peak load with 5x the nominal current for 5ms, typ. 10ms/div output voltage output current 24v 0a 20a 17v 1ms/div output voltage output current 24v 0a 50a 15.5v 20a peak load (resistive) for 50ms output voltage dips from 24v to 17v. 50a peak load (resistive) for 5ms output voltage dips from 24v to 15.5v. fig. 24-3 30a peak load, typ. 10ms/div output voltage output current 24v 0a 12ms 30a 10.5v please note: the dc-ok relay triggers when the voltage dips more than 10% for longer than 1ms. high overload current (typ. 30a for 12ms) enables easy fuse tripping peak current voltage dips typ. from 24v to 17v at 20a for 50ms, resistive load typ. from 24v to 19v at 50a for 2ms, resistive load typ. from 24v to 15.5v at 50a for 5ms, resistive load may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 23/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 24.2. b ack - feeding l oads loads such as decelerating motors and induc tors can feed voltage back to the power supply. this feature is also called return voltage immunity or resi stance against back- e.m.f. ( e lectro m agnetic f orce). this power supply is resistant and does not show malfunctioning when a load f eeds back voltage to the power supply. it does not matter whether the power supply is on or off. the maximum allowed feed-b ack-voltage is 35vdc. the maximum allowed feed-back peak current is 40a. higher currents can temporarily shut-down the ou tput voltage. the absorbing energy can be calculated according to the built-in large sized output capacito r which is specified in chapter 6 . 24.3. e xternal i nput p rotection the unit is tested and approved for bran ch circuits up to 30a (ul) and 32a (i ec). an external protection is only required if the supplying branch has an ampacity greater than this. check also local codes and local requirements. in some countries local re gulations might apply. if an external fuse is necessary or utilized, minimum requirements need to be considered to avoid nuisance tripping of the circuit breaker. a mi nimum value of 6a b- or c-chara cteristic breaker should be used. 24.4. o utput c ircuit b reakers standard miniature circuit breakers (mcbs or ul 1077 circuit breakers) are commonly used for ac-supply systems and may also be used on 24v branches. mcbs are designed to protect wires and ci rcuits. if the ampere value and the char acteristics of the mcb are adapted to the wire size that is used, the wiring is considered as thermally safe rega rdless of whether the mcb opens or not. to avoid voltage dips and under-voltage situations in adjacent 24v branches wh ich are supplied by the same source, a fast (magnetic) tripping of the mcb is desired. a quick shutdown within 10ms is necessary corresponding roughly to the ride-through time of plc's. this re quires power supplies with high current reserves and large output capacitors. furthermore, the impedance of the faulty branch must be sufficiently small in or der for the current to actually flow. the best current reserve in the power su pply does not help if ohms law does not permit current flow. the following table has typical test results showing which b- and c-characteristic mcbs magneti cally trip depending on the wire cross section and wire length. fig. 24-4 test circuit maximal wire length *) for a fast (magnetic) tripping: 0.75mm2 1.0mm2 1.5mm2 2.5mm2 c-2a 30m 37m 54m 84m c-3a 25m 30m 46m 69m c-4a 9m 15m 25m 34m c-6a 3m 3m 4m 7m b-6a 12m 15m 21m 34m b-10a 3m 3m 4m 9m b-13a 2m 2m 3m 6m mcb power supply ac dc + - + - load wire length s1... fault simulation switch s1 *) dont forget to consider twice the distance to the load (or cable length) when calculating the total wire length (+ and C wi re). may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 24/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 24.5. s eries o peration do not use the power supply in series . the leakage current will be too high to meet the medical requirements. 24.6. p arallel u se to i ncrease o utput p ower do not use the power supply in parallel. the leakage current will be too high to meet the medical requirements. 24.7. p arallel u se for r edundancy do not use the power supply in parallel. the leakage current will be too high to meet the medical requirements. 24.8. i nductive and c apacitive l oads the unit is designed to supply any kind of loads, including capacitive and inductive loads. if extreme large capacitors, such as edlcs (electric double layer capacitors or ultracaps) with a capaci tance larger than 1.5f are connected to the output, the unit might charge the capacitor in the hiccup plus mode (see chapter 6 ). 24.9. c harging of b atteries the power supply can be used to charge lead-acid or maintenanc e free batteries (sla or vrla batteries). two 12v batteries are needed in series. instructions for charging batteries: a) set output voltage (measured at no load and at the battery end of the cable) very precisely to the end-of-charge voltage. end-of-charge voltage 27. 8v 27.5v 27.15v 26.8v battery temperature 10c 20c 30c 40c b) use a 15a or 16a circuit breaker (or blocking diode) between th e power supply and the battery. c) ensure that the output current of the power supply is below the allowe d charging current of the battery. d) use only matched batteries when putting 12v types in series. e) ensure that the ambient temperature of the power supply stays below 40c. f) the return current to the power supply (battery discharge current is ty p. 3.5ma when the power supply is switched off (except in case a blocking diode is utilized). may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 25/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput 24.10. o peration on t wo p hases 240v +10% max. l2 l1 l3 l n pe power supply ac dc the power supply can also be used on two-phases of a three-phase- system. such a phase-to-phase connection is allowed as long as the supplying voltage is below 240v +10% . 24.11. u se in a t ightly s ealed e nclosure when the power supply is installed in a tightly sealed encl osure, the temperature inside the enclosure will be higher than outside. in such situations, th e inside temperature defi nes the ambient temperatur e for the power supply. the following measurement results can be used as a reference to estimate the temperature rise inside the enclosure. the power supply is placed in the middle of the box , no other heat producing items are inside the box input: 230vac case a: enclosure: rittal typ ip66 box pk 9519 100, plastic, 180x180x165mm load: 24v, 8a; (=80%) load is placed outside the box temperature inside the box: 42.0c (in the middle of the right side of the pow er supply with a distance of 1cm) temperature outside the box: 25.8c temperature rise: 16.2k case b: enclosure: rittal typ ip66 box pk 9519 100, plastic, 180x180x165mm load: 24v, 10a; load is placed outside the box temperature inside the box: 48.1c (in the middle of the right side of the pow er supply with a distance of 1cm) temperature outside the box: 26.2c temperature rise: 21.9k case c: enclosure: rittal typ ip66 box pk 9516 100, plastic, 110x180x165mm load: 24v, 8a; (=80%) load is placed outside the box temperature inside the box: 48.6c (in the middle of the right side of the pow er supply with a distance of 1cm) temperature outside the box: 26.3c temperature rise: 22.3k case d: enclosure: rittal typ ip66 box pk 9519 100, plastic, 110x180x165mm load: 24v, 10a; load is placed outside the box temperature inside the box: 53.8c (in the middle of the right side of the pow er supply with a distance of 1cm) temperature outside the box: 26.6c temperature rise: 27.3k may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 26/27
cp10.241-m1 cp-series 24v, 10a, s ingle p hase i nput may 2017 / rev. 1.2 ds-cp10.241-m1-en all parameters are specified at 24v, 10a, 230vac, 50hz, 25c ambient and after a 5 minutes run-in time unless otherwise noted. www.pulspower.com phone +49 89 9278 0 germany 24.12. m ounting o rientations mounting orientations ot her than all terminals on the bottom require a reduction in continuous output power or a limitation in the maximum allowed ambien t temperature. the amount of reduction influences the lifetime expectancy of the power supply. therefore, tw o different derating curves for cont inuous operation can be found below: curve a1 recommended output current. curve a2 max allowed output current (res ults in approximately half the lifetime expectancy of a1). fig. 24-5 mounting orientation a (standard orientation) power supply output input output current 0 10 20 30 40 60 3 6 9 12a 50 a 1 70c ambient temperature 10a 7.5a fig. 24-6 mounting orientation b (upside down) power supply output input output current 0 10 20 30 40 60 3 6 9 12a 50 a 2 a 1 70c ambient temperature 6.8a 5.6a fig. 24-7 mounting orientation c (table-top mounting) output current 0 10 20 30 40 60 3 6 9 12a 50 a 1 a 2 70c 5.8a 4 . 3 a ambient temperature fig. 24-8 mounting orientation d (horizontal cw) power supply output input output current 0 10 20 30 40 60 3 6 9 12a 50 a 1 a 2 70c ambient temperature 4.5a 5 . 7 a fig. 24-9 mounting orientation e (horizontal ccw) power supply output input output current 0 10 20 30 40 60 3 6 9 12a 50 a 1 a 2 70c ambient temperature 5 . 2 a 4 . 2 a 27/27


▲Up To Search▲   

 
Price & Availability of CP10241-M1

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X